Yokogawa MG8E-5AWNS-E Flameproof Paramagnetic Oxygen Analyzer
Continuous oxygen measurement in flammable process gas with nitrogen auxiliary gas
Equipment Profile
Selective oxygen measurement in combustible process mixtures is the defining purpose of Yokogawa MG8E-5AWNS-E. The analyzer uses the paramagnetic property of oxygen and a magnetic proportional-flow cell, allowing oxygen concentration to be measured without placing a moving sensor element in the sample stream.
Inside MG8E-5AWNS-E, clean nitrogen auxiliary gas is divided into two streams around the magnetic cell. Oxygen in the sample changes the flow balance between those streams; thermistors detect that difference and the electronics convert it into the displayed concentration and isolated 4-20 mA output.
The configured MG8E-5AWNS-E record covers a 0-5 to 25 vol% oxygen range, standard cell material, nitrogen auxiliary gas and a 35 ml/min auxiliary-flow arrangement. Its flameproof Ex d IIB T4 construction, 100-115 VAC supply and English interface belong to the complete order identity.
Operating Data
| Technical Item | MG8E-5AWNS-E |
|---|---|
| Brand | Yokogawa |
| Product Type | Flameproof paramagnetic oxygen analyzer |
| Brand Country | Japan |
| Measurement Object | Oxygen concentration in gaseous mixtures |
| Measurement Range | 0-5 to 25 vol% O2 |
| Measurement System | Magnetic proportional-flow paramagnetic system |
| Analog Output | Isolated 4-20 mA DC; 550 ohm maximum load |
| Response Time | 90% response within 3 seconds |
| Repeatability / Linearity | Within +/-1% of span |
| Sample Gas Flow | 300-800 ml/min; 600 ml/min standard |
| Analyzer Inlet Pressure | Approximately 7 kPa |
| Sample Temperature | 0 to 50 deg C |
| Auxiliary Gas | Nitrogen, 350-500 kPa; 35 ml/min |
| Power / Construction | 100-115 VAC, 50/60 Hz; flameproof Ex d IIB T4 |
Magnetic Proportional-Flow Measurement
A balanced auxiliary-gas path converts oxygen attraction into an electrical signal.
MG8E-5AWNS-E does not expose the measuring thermistors directly to the process gas. Two auxiliary-gas streams meet the sample inside the ring-shaped sensor cell, and oxygen drawn toward the magnetic field changes one stream relative to the other.
Thermistor bridge
The paired thermistors in MG8E-5AWNS-E sense the differential flow instead of measuring oxygen by a consumable electrochemical reaction.
Background compensation
Cell-angle adjustment in MG8E-5AWNS-E compensates the density-related influence of a defined background gas or gas mixture.
Sample, Auxiliary-Gas and Calibration Paths
A regulated 300-800 ml/min sample enters MG8E-5AWNS-E at approximately 7 kPa and leaves through the common gas outlet.
Clean nitrogen at 350-500 kPa feeds the protected sensor-cell path of MG8E-5AWNS-E at the configured 35 ml/min flow.
Zero gas and span gas can be switched through MG8E-5AWNS-E for manual, semiautomatic or programmed automatic calibration.
Purchasing Questions
Why should nitrogen flow be established before sample gas reaches MG8E-5AWNS-E?
The protected measuring cell of MG8E-5AWNS-E depends on clean nitrogen auxiliary gas to isolate the thermistors from the process mixture and establish the balanced flow reference. Loss of that flow is therefore an operating fault, not a minor flow variation.
What sample-flow window keeps MG8E-5AWNS-E inside its stated operating conditions?
MG8E-5AWNS-E is specified for 300-800 ml/min, with 600 ml/min as the standard setting. The inlet pressure is approximately 7 kPa, so the sampling system must regulate both flow and pressure rather than forcing plant pressure into the analyzer.
Can MG8E-5AWNS-E measure oxygen when the process mixture is flammable?
Yes. MG8E-5AWNS-E is the flameproof MG8E construction and is identified as Ex d IIB T4. The installed wiring, cable entries and sampling system must still preserve the certified enclosure and site classification.
How quickly can MG8E-5AWNS-E show a changing oxygen concentration?
The documented response of MG8E-5AWNS-E reaches 90% within three seconds at the analyzer. Sampling-line volume, filters and conditioning equipment can add transport delay before the gas reaches the cell.
Which calibration gases are appropriate for MG8E-5AWNS-E?
MG8E-5AWNS-E uses nitrogen as zero gas and a known oxygen-in-nitrogen mixture, or dry instrument air where appropriate, as span gas. The entered concentration must match the cylinder value used for the calibration.
What does the 0-5 to 25 vol% range statement mean for MG8E-5AWNS-E?
MG8E-5AWNS-E supports configured oxygen ranges whose span lies from 5 to 25 vol% O2. It is not a 0-100% analyzer record, so the selected output range must stay within that order-code boundary.
Analyzer Documentation Desk
Moore Automation
https://www.mvme.cn/Moore Automation keeps analyzer identity, available technical documents and destination records aligned for Yokogawa MG8E-5AWNS-E. Specialized process-instrumentation requests follow a single communication chain, preserving the full configuration while documents and logistics are coordinated.
Selected Products
| Control & Process | PLC & I/O | Monitoring & Network | Drives & Power |
|---|---|---|---|
| 01Emerson PR9268-301-100 | 16Foxboro FBM214B | 31Siemens 6ES7153-2BA10-0XB0 | 46ABB AI810 |
| 02ABB CI871K01 | 17ICS Triplex T8830C | 32GE IC630CCM390 | 47Siemens 6ES7321-1BL00-0AA0 |
| 03Yokogawa ACCC01 | 18Allen-Bradley 22B-A1P5N104 | 33Horner HE800THM000 | 48GE IC630MDL302 |
| 04Siemens 6ES7407-0DA02-0AA0 | 19Emerson PR6424-003-030 | 34Foxboro RH101AB | 49Horner HE800MIX902 |
| 05GE IC630CCM300 | 20ABB TB850 | 35ICS Triplex T8312-7C | 50Foxboro RH102AG |
| 06Horner HE800DAC106 | 21Yokogawa AKB337 | 36Allen-Bradley 22B-A3P6N104 | 51ICS Triplex T8151B |
| 07Foxboro RH916XG | 22Siemens 6ES7407-0KR02-0AA0 | 37Emerson PR9268-202-100 | 52Allen-Bradley 22B-A8P0N104 |
| 08ICS Triplex T8800C | 23GE IC630CCM311 | 38ABB TB852 | 53Emerson PR6423-010-000 |
| 09Allen-Bradley 22B-D030N104 | 24Horner HE800RTD100 | 39Siemens 6ES7153-4BA00-0XB0 | 54ABB AI815 |
| 10Emerson PR9268-302-100 | 25Foxboro RH927AH | 40GE IC630MDL301 | 55Siemens 6ES7321-7BH01-0AB0 |
| 11ABB CI873K01 | 26ICS Triplex T8850C | 41Horner HE800THM100 | 56GE IC630MDL303 |
| 12Yokogawa AKB331 | 27Allen-Bradley 22B-A2P3N104 | 42Foxboro RH924WA | 57Horner HE800ACM200 |
| 13Siemens 6ES7407-0KA02-0AA0 | 28Emerson A6120 | 43ICS Triplex T8110B | 58Foxboro FBM217 |
| 14GE IC630CCM310 | 29ABB TB851 | 44Allen-Bradley 22B-A5P0N104 | 59ICS Triplex T8403 |
| 15Horner HE800RTD000 | 30Yokogawa KS1 | 45Emerson CON041 | 60Allen-Bradley 22B-B2P3N104 |









No related record found