Heated elements
Platinum wires operate as temperature-sensitive resistive elements inside the two gas chambers.
Welcome to Moore Automation Limited
Copy and share this link on social network or send it to your friends
Copy| Product Name | Fuji Electric ZAF3K404-MYAAA-YEAYY Thermal Conductivity Gas Analyzer |
| Item NO. | 2944911 |
| Weight | 0.1 kg = 0.2205 lb = 3.5274 oz |
| Category | FUJI |
| Brand | FUJI |
| Creation Time | 2026-11-25 |
Fuji Electric ZAF3K404-MYAAA-YEAYY is a panel-mounted thermal conductivity gas analyzer. Heated platinum wires in sample and reference chambers form a Wheatstone bridge whose imbalance changes with the thermal conductivity of the gas mixture.
Fuji Electric ZAF3K404-MYAAA-YEAYY supports concentration measurement for gases including helium, argon, hydrogen, methane and carbon dioxide when the measuring and reference-gas composition suits the thermal-conductivity method. The front display and keys handle operating values, zero/span work and settings.
Fuji Electric ZAF3K404-MYAAA-YEAYY routes sample gas through rear inlet and outlet ports while the rear terminal area carries power, measured output and optional contact/communication functions. Stable flow, clean dry tubing and atmospheric-pressure exhaust preserve the bridge response.
| Specification | ZAF3K404-MYAAA-YEAYY |
|---|---|
| Brand | Fuji Electric |
| Product Type | Thermal conductivity gas analyzer |
| Brand Country | Japan |
| Measuring principle | Thermal conductivity / Wheatstone bridge |
| Supported gas components | He, Ar, H2, CH4, CO2 |
| Repeatability | ±1% of full scale |
| Linearity | ±2% of full scale |
| Standard response time | 60 s or less at 0.4 L/min |
| High-speed response | 10 s or less at 1 L/min for H2 |
| Standard sample flow | 0.4 ±0.05 L/min |
| High-speed sample flow | 1.0 ±0.05 L/min |
| Analog output options | 4–20 mA, 0–1 V or 0–10 mV |
| Optional communication | RS-232C Modbus |
| Power supply | 100–240 V AC, 50/60 Hz |
| Power consumption | Approximately 50 VA |
| Warm-up time | 30 minutes or more |
| Operating temperature | -5 to 45 °C |
| Storage temperature | -20 to 60 °C |
| Humidity | 90% RH or less, non-condensing |
| Approximate mass | 5 kg |
Fuji Electric ZAF3K404-MYAAA-YEAYY converts the heat-transfer difference between sample and reference gas into an electrical concentration signal.
Platinum wires operate as temperature-sensitive resistive elements inside the two gas chambers.
Reference and sample elements form a Wheatstone bridge that balances at the established zero condition.
A sample-gas conductivity change alters wire temperature, unbalances the bridge and produces the measured output.
Fuji Electric ZAF3K404-MYAAA-YEAYY depends on a controlled gas path because flow, moisture and outlet pressure directly affect heat transfer around the measuring elements.
The standard path uses 0.4 L/min; the documented high-speed H2 response uses 1 L/min.
Final filtration and a low dew point prevent dust or condensation from coating the measuring chamber.
The outlet is relieved to atmosphere so backpressure does not distort flow or strain internal piping.
Fuji Electric ZAF3K404-MYAAA-YEAYY needs at least 30 minutes for the heated bridge and internal temperature to stabilize; adjustment during warm-up can create a drifting baseline.
Fuji Electric ZAF3K404-MYAAA-YEAYY uses 0.4 ±0.05 L/min in the standard configuration. The 1 L/min condition belongs to the high-speed hydrogen response arrangement.
Condensation inside Fuji Electric ZAF3K404-MYAAA-YEAYY changes heat transfer and may contaminate the chamber. Keep the sample clean, filtered and near the documented dew-point condition before it reaches the inlet.
Long tubing, excessive internal volume, low flow or a restricted outlet delay gas exchange at Fuji Electric ZAF3K404-MYAAA-YEAYY. Inspect the entire sample path before changing analyzer settings.
Fuji Electric ZAF3K404-MYAAA-YEAYY supports ordered analog-output forms including 4–20 mA, 0–1 V and 0–10 mV, with optional contact I/O and RS-232C Modbus functions.
Apply the correct zero or span gas to Fuji Electric ZAF3K404-MYAAA-YEAYY at the same controlled flow used for measurement, wait for a stable reading and keep the outlet at atmospheric pressure.