Product Name Fuji Electric ZAF3K404-MYAAA-YEAYY Thermal Conductivity Gas Analyzer
Item NO. 2944911
Weight 0.1 kg = 0.2205 lb = 3.5274 oz
Category FUJI
Brand FUJI
Creation Time 2026-11-25

Fuji Electric ZAF3K404-MYAAA-YEAYY Thermal Conductivity Gas Analyzer
Fuji ElectricZAFThermal Conductivity Gas Analyzer
Panel-mounted thermal-conductivity measurement

Fuji Electric ZAF3K404-MYAAA-YEAYY Thermal Conductivity Gas Analyzer

Measuring principleThermal conductivity / Wheatstone bridge
Supported gas componentsHe, Ar, H2, CH4, CO2
Repeatability±1% of full scale
01 / PRODUCT FACTS

Product Description

Fuji Electric ZAF3K404-MYAAA-YEAYY is a panel-mounted thermal conductivity gas analyzer. Heated platinum wires in sample and reference chambers form a Wheatstone bridge whose imbalance changes with the thermal conductivity of the gas mixture.

Fuji Electric ZAF3K404-MYAAA-YEAYY supports concentration measurement for gases including helium, argon, hydrogen, methane and carbon dioxide when the measuring and reference-gas composition suits the thermal-conductivity method. The front display and keys handle operating values, zero/span work and settings.

Fuji Electric ZAF3K404-MYAAA-YEAYY routes sample gas through rear inlet and outlet ports while the rear terminal area carries power, measured output and optional contact/communication functions. Stable flow, clean dry tubing and atmospheric-pressure exhaust preserve the bridge response.

02 / VERIFIED DATA

Technical Data

Specification ZAF3K404-MYAAA-YEAYY
Brand Fuji Electric
Product Type Thermal conductivity gas analyzer
Brand Country Japan
Measuring principle Thermal conductivity / Wheatstone bridge
Supported gas components He, Ar, H2, CH4, CO2
Repeatability ±1% of full scale
Linearity ±2% of full scale
Standard response time 60 s or less at 0.4 L/min
High-speed response 10 s or less at 1 L/min for H2
Standard sample flow 0.4 ±0.05 L/min
High-speed sample flow 1.0 ±0.05 L/min
Analog output options 4–20 mA, 0–1 V or 0–10 mV
Optional communication RS-232C Modbus
Power supply 100–240 V AC, 50/60 Hz
Power consumption Approximately 50 VA
Warm-up time 30 minutes or more
Operating temperature -5 to 45 °C
Storage temperature -20 to 60 °C
Humidity 90% RH or less, non-condensing
Approximate mass 5 kg
03 / FUNCTIONAL VIEW

Thermal-Conductivity Bridge

ZAF3K404-MYAAA-YEAYY PROCESS

Thermal-Conductivity Bridge

Fuji Electric ZAF3K404-MYAAA-YEAYY converts the heat-transfer difference between sample and reference gas into an electrical concentration signal.

01

Heated elements

Platinum wires operate as temperature-sensitive resistive elements inside the two gas chambers.

02

Bridge balance

Reference and sample elements form a Wheatstone bridge that balances at the established zero condition.

03

Concentration signal

A sample-gas conductivity change alters wire temperature, unbalances the bridge and produces the measured output.

04 / SYSTEM ARRANGEMENT

Sample Gas Conditioning

ZAF3K404-MYAAA-YEAYY ARRANGEMENT

Sample Gas Conditioning

Fuji Electric ZAF3K404-MYAAA-YEAYY depends on a controlled gas path because flow, moisture and outlet pressure directly affect heat transfer around the measuring elements.

01

Flow control

The standard path uses 0.4 L/min; the documented high-speed H2 response uses 1 L/min.

02

Clean dry sample

Final filtration and a low dew point prevent dust or condensation from coating the measuring chamber.

03

Free exhaust

The outlet is relieved to atmosphere so backpressure does not distort flow or strain internal piping.

05 / PRACTICAL CHECKS

Product Q&A

Q01

Why should Fuji Electric ZAF3K404-MYAAA-YEAYY warm up before zero and span work?

ANSWER

Fuji Electric ZAF3K404-MYAAA-YEAYY needs at least 30 minutes for the heated bridge and internal temperature to stabilize; adjustment during warm-up can create a drifting baseline.

Q02

What sample flow is expected by Fuji Electric ZAF3K404-MYAAA-YEAYY in standard operation?

ANSWER

Fuji Electric ZAF3K404-MYAAA-YEAYY uses 0.4 ±0.05 L/min in the standard configuration. The 1 L/min condition belongs to the high-speed hydrogen response arrangement.

Q03

How does moisture affect Fuji Electric ZAF3K404-MYAAA-YEAYY?

ANSWER

Condensation inside Fuji Electric ZAF3K404-MYAAA-YEAYY changes heat transfer and may contaminate the chamber. Keep the sample clean, filtered and near the documented dew-point condition before it reaches the inlet.

Q04

What can make Fuji Electric ZAF3K404-MYAAA-YEAYY respond slowly even when the display is normal?

ANSWER

Long tubing, excessive internal volume, low flow or a restricted outlet delay gas exchange at Fuji Electric ZAF3K404-MYAAA-YEAYY. Inspect the entire sample path before changing analyzer settings.

Q05

Which signal forms can Fuji Electric ZAF3K404-MYAAA-YEAYY provide to another system?

ANSWER

Fuji Electric ZAF3K404-MYAAA-YEAYY supports ordered analog-output forms including 4–20 mA, 0–1 V and 0–10 mV, with optional contact I/O and RS-232C Modbus functions.

Q06

How should zero and span gases be introduced to Fuji Electric ZAF3K404-MYAAA-YEAYY?

ANSWER

Apply the correct zero or span gas to Fuji Electric ZAF3K404-MYAAA-YEAYY at the same controlled flow used for measurement, wait for a stable reading and keep the outlet at atmospheric pressure.

06 / MOORE AUTOMATION

Extensive Inventory

Moore Automation

https://www.mvme.cn/

Moore Automation maintains a wide range of industrial automation components for diverse applications. The inventory spans PLC and DCS platforms, operator interfaces, industrial computers, drive and power products, sensors, communication modules and supporting accessories.

PLC and DCSController and control-system components
HMI and IPCOperator and industrial-computing products
Drives and powerMotion and power-conversion families
Sensors and modulesField devices and supporting I/O
07 / MIXED BRAND CATALOG

Recommended Products

CATALOG ACATALOG BCATALOG CCATALOG D
Siemens6ES7654-7HY00-0XA0
GEIC630PWR310
HornerHE800DIQ722
FoxboroP0914TN
ICS TriplexT8290C
Allen-Bradley1756-PA72
EmersonVE4005S2B5
ABBCI858K01
YokogawaARM15B
Siemens6ES7654-7HX00-0XA0
GEIC630PWR314
HornerHE800ADC010
FoxboroRH914TN
ICS TriplexT8431C
Allen-Bradley1756-PA75
EmersonCE281
ABBCI860K01
YokogawaARS15M
Siemens6ES7654-7HB00-0XA0
GEIC630PWR320
HornerHE800ADC020
FoxboroRH916BT
ICS TriplexT8451C
Allen-Bradley1756-PB72
EmersonCE620
ABBCI862K01
YokogawaARM55D
Siemens6ES7654-7HC00-0XA0
GEIC630PWR324
HornerHE800ADC110
FoxboroRH916JT
ICS TriplexT8311C
Allen-Bradley1756-PB75
EmersonCON021
ABBCI865K01
YokogawaARM55W
Siemens6ES7400-1JA11-0AA0
GEIC630CHS304
HornerHE800ADC120
FoxboroFBM201
ICS TriplexT8311CX
Allen-Bradley25C-D017N104
EmersonPR6423-002-030
ABBCI867K01
YokogawaARM55C
Siemens6ES7400-1TA11-0AA0
GEIC630CHS306
HornerHE800ADC140
FoxboroP0914SQ
ICS TriplexT8100C
Allen-Bradley25C-D024N104
EmersonPR9268-201-100
ABBCI868K01
YokogawaARS55M
Siemens6ES7405-0DA02-0AA0
GEIC630CHS308
HornerHE800ADC920
FoxboroRH914SQ
ICS TriplexT8300C
Allen-Bradley25C-D030N104

No related record found

You May Also Like

Copyright © 2004–2026 www.mvme.cn. All rights reserved.

Disclaimer: Moore Automation Limited is an independent reseller and is not an authorized distributor or representative of the manufacturers featured on this website. Product names and trademarks belong to their respective owners.